The AirSense 10 Elite is an advanced fixed-pressure therapy device with an integrated humidifier. It is designed to be intuitive and easy-to-use. With no complicated menus or settings to navigate, patients can simply plug in their device and press Start.
Current AirSense 10 devices do not have built-in cellular connectivity functionality. This means the device has no wireless data transmission capability with myAir, ResMed's online support program and app.
The AirSense 10 Elite can detect indications of central sleep apnea and Cheyne-Stokes respiration. By differentiating between different types of sleep disordered breathing, you can be confident that your patients are receiving appropriate therapy.
ResMed’s AutoRamp™ feature with sleep onset detection starts by delivering a low pressure for greater patient comfort. Then, once the patient falls asleep, it increases the pressure to the prescribed level.
AirSense 10 Elite
Nasal Pillows Mask.
Nasal Pillows Mask in a stylish, female design.
The ClimateLineAir heated tube is the key that unlocks ResMed's Climate Control solution for the AirSense 10 and AirCurve 10 series of devices.
The SlimLine small diameter air tube is exceptionally light and flexible, which can reduce tube drag.
The ClimateLine heated tube is the key technology in the Climate Control system.
Designed to seamlessly integrate with the device, the HumidAir heated humidifier offers relief from dryness and congestion.
CPAP | Elite | AutoSet | AutoSet for Her | |
---|---|---|---|---|
CPAP | ✓ | ✓ | ✓ | ✓ |
AutoSet | ✓ | ✓ | ||
AutoSet for Her | ✓ | |||
Cheyne-Stokes respiration detection | ✓ | ✓ | ✓ | |
Central apnea detection | ✓ | ✓ | ✓ | |
RERA reporting | ✓ | |||
AutoSet algorithm with AutoSet Response | ✓ | ✓ | ||
AutoRamp with sleep onset detection | ✓ | ✓ | ✓ | ✓ |
Advanced HumidAir technology | ✓ | ✓ | ✓ | ✓ |
Enhanced Climate Control Auto | ✓ | ✓ | ✓ | ✓ |
Expiratory Pressure Relief (EPR™) | ✓ | ✓ | ✓ | ✓ |
SmartStart™ | ✓ | ✓ | ✓ | ✓ |
Built-in wireless connectivity* | ✓ | ✓ | ✓ | ✓ |
Dimensions (H x W x D) (Device and water tub) | 116mm x 255mm x 150mm | ||
---|---|---|---|
Weight | 1248g | ||
Air filter | Standard: Average arrestance: > 75% for ~7 micron dust | Hypoallergenic: Efficiency: >98% for ~7-8 micron dust. >80% for ~0.5 micron dust. | |
Aircraft use | Meets Federal Aviation Administration requirements for all phases of air travel. | ||
Operating pressure range | 4 to 20 cm H2O | ||
Pressure level measured according to ISO 17510-1:2007 (CPAP mode): | |||
SlimLine or Standard air tubing | 26.6 dBA with uncertainty of 2 dBA | ||
Power level measured according to ISO 17510-1:2007 (CPAP mode): | |||
SlimLine or Standard air tubing | 34.6 dBA with uncertainty of 2 dBA | ||
Sound (Declared dual-number noise emission values in accordance with ISO 4871:1996.) | |||
Air tubing | ClimateLineTM Air | SlimLineTM | Standard |
Length | 6'6” | 6’ | 6’ 6” |
Inner diameter | 0.6" | 0.6" | 0.75" |
Electromagnetic compatibility | Complies with all EMC requirements according to IEC60601-1-2:2007 for residential, commercial and light industry environments. It is recommended that mobile communication devices be kept at least 1m away. |